Galvo Add-On - Scanning Microscopy Interface
The Galvo Box (mAIkroscope Galvo Interface Board) brings advanced scanning capabilities to UC2 microscopes through precision galvanometer mirror control. This system enables laser scanning microscopy, time-resolved imaging, and synchronization with external detectors.
Together with FLIM LABS it forms the FLIM laser scanner for the openUC2 FRAME (module
OPM - 0004 - FLIM 488). A 488 nm fibre laser is scanned by a galvo pair through an f50 /
f160 telescope into the objective. The fluorescence returns the same way to a detection
fibre that acts as a confocal pinhole.
Where do I start?
These pages follow Diátaxis: four kinds of page for four needs.
I am building or aligning the scanner => Tutorial
- Build and align the FLIM laser scanner: eleven steps from the fibre to a scanned image, each with a picture and a check.
Something specific needs fixing => How-to guides
- Re-align after transport: only two kinematic cubes and the galvo centre.
- Fix a clipped or off-centre beam: D-shaped, crescent or missing spot at the objective.
- Fix a stretched image: one axis longer than the other.
- Choose an objective: 4x, 10x or 20x, and how to measure the PSF.
- Stripes, dots or lines in FLIM images: from what you see to what to do.
- A second peak in the FLIM decay: choose and check the detector and sync cables.
I want to understand it => Explanation
- How laser scanning works: confocal pinhole, 4f pupil relay, filling the pupil, descanned detection, timing.
- Optical quality and limits: what an Optiland simulation of the module says about clipping, field curvature, the galvo mirrors and alignment errors.
- Stripes, dots and lines in FLIM images: what the scan configuration, the USB link and the pixel clock timing do to a FLIM LABS image.
- Signal cables and a second peak in the decay: an echo on the detector cable, how the bench tests found it, and why to calibrate again after any cable change.
I need a number => Reference
- Optical layout of the FLIM module
- Scan parameters: every field of the raster command, what it does to the image, and the recommended values for FLIM
- Optical simulation (Optiland)
- Galvo interface board
- Trigger outputs (measured)
- Galvo driver board: the red ±15 V servo drivers, their trimmers and how to identify them
What is Galvo Scanning?
Galvanometer mirrors are electrically controlled mirrors that can rapidly and precisely steer laser beams in 2D. By scanning a focused laser beam across a sample and synchronizing detection with the beam position, you can create high-resolution images or perform point-by-point measurements.
Key Features
Dual-Axis Galvo Control
- On-board 12-bit SPI DAC generates ±10V differential XY signals
- Compatible with low-cost Chinese galvo drivers
- Seeed XIAO ESP32-S3 MCU for scan synthesis
Synchronized Triggering
- Three independent 50Ω trigger outputs (pixel, line, frame)
- U.FL laser blanking output
- Precise timing for detector synchronization
CAN Bus Integration
- CAN 2.0B bus for remote control
- Compatible with UC2-CAN protocol
- Network multiple devices together
Power System
- Integrated ±12V/3.3V/5V analog rails
- 2A output budget for external driver modules
- Complete power solution for galvo systems
What You'll Learn
- Galvanometer mirror control and calibration
- Laser scanning microscopy principles
- DAC (Digital-to-Analog Converter) operation
- Trigger signal generation and synchronization
- CAN bus communication protocols
- High-speed image acquisition
- Integration with time-resolved detectors
Applications
- Laser scanning microscopy
- Two-photon excitation microscopy
- Fluorescence lifetime imaging (FLIM)
- Point-scanning confocal microscopy
- Optogenetics with spatial control
- Material ablation and lithography
- Synchronized multi-modal imaging
Technical Specifications
| Parameter | Value | Notes |
|---|---|---|
| Backbone input | 12V ±5% | via JST connector |
| Board consumption | ~50mA | excludes external drivers |
| Galvo supply | ±12V, 2A max | pass-through to drivers |
| DAC resolution | 12-bit | 4096 steps over ±10V |
| Output bandwidth | ~25kHz (-3dB) | limited by amplifier stage |
| Trigger level | 3.3V CMOS | 50Ω back-terminated |
| CAN bus | ISO11898-2, 1Mbit/s | 120Ω termination selectable |
System Architecture
ESP32-S3 ↔ SPI ↔ MCP4822 DAC → Differential Amps → Galvo Drivers
│ │
│ ├─ Pixel/Line/Frame triggers
│ └─ Laser blanking
└─ CAN bus → External control
Required Components
- mAIkroscope Galvo Interface Board
- Dual-axis galvanometer mirror system
- External galvo drivers (±10V input)
- Laser source with blanking input
- Detector with trigger input (e.g., FLIM detector)
- 12V power supply
- UC2 mounting components
Integration Examples
Successfully integrated with:
- FLIMlabs.com detector systems
- Pulsed laser sources
- Time-resolved photon counting systems
- Custom UC2-CAN networks
Perfect for researchers in advanced microscopy, biophysics, and anyone interested in high-speed laser scanning and time-resolved imaging techniques!