Skip to main content

How to re-align the scanner after transport

Task-oriented. Use this when the system worked before and now gives a weak, cut or off-centre beam.

The lenses and the galvo sit in cubes on the 50 mm grid and do not drift. After transport you touch two kinematic cubes and one software setting:

whatmoveshow you see it
junction dichroic (next to the excitation fibre)the excitation beam's direction into the galvo, so its position on the galvo mirrors and in the objective pupilspot position on the mirrors; overlap with the red detection laser far away
splitter under the objective (end of the scan row)the beam's direction into the objective, so the field centre and the beam position in the pupilspot on a card at the objective position
galvo centre (ImSwitch, DAC X / Y)the scan centreposition of the scan field in the camera image

Before you start​

  • Goggles on, laser at minimum power, objective removed.
  • Galvo powered and parked at your centre values (from the tutorial, step 5; default 2048 / 2048).

Step 1 - Beam on the galvo mirror centres​

Look onto the two galvo mirrors from above through the open galvo cube.

  • Centred on both: go to step 2.
  • Off-centre, especially across the short (6 mm) side: correct it with the junction dichroic. Then check that the excitation still overlaps the red detection laser far away (tutorial step 3, the "far" part). If not, re-do that step.

Step 2 - Beam at the objective position​

Put the module into the FRAME (objective still out) and place a card on the objective changer.

  • Round and centred on the 14 mm opening: go to step 3.
  • Off-centre: centre it with the splitter's tip / tilt screws, in small steps.
  • D-shaped, crescent or missing: work through Fix a clipped or off-centre beam first.

Step 3 - Scan field in the image​

Objective in, fluorescent slide in focus, widefield camera on.

  • Send the beam to the centre and look where the spot is in the camera image.
  • Small offsets (up to about 1° optical, which is about 250 µm with the 4x): use the galvo centre in software.
  • Larger offsets: the splitter is tilted. Correct it there (step 2), not with the galvo: a large galvo offset pushes the beam towards the edges of the lenses and the pupil.
What each adjustment costs

From the simulation, with the 4x and a 2 mm beam:

  • A 0.5° tilt of the splitter turns the beam by 1°. That moves the spot about 800 µm in the sample and moves the beam 1.8 mm across the objective pupil.
  • A 2° optical offset of the galvo centre costs little at the centre (8 %), but about two thirds of the peak at one edge of the field.

Details: Optical quality and limits.

Step 4 - Peak up the detection​

With a bright uniform sample, scan a small area and maximise the mean count rate with the detection fibre holder's XY (tutorial step 11).

Don't​

  • Don't move the scan lens, the tube lens or the galvo cube. They are on the grid and are not the cause.
  • Don't adjust both kinematic cubes at once: change one, check, then the next.
  • Don't use the galvo offset to compensate a beam that sits on a mirror edge.