openSIM with the DLP2000
Status: rev B designed; next are printing and the first optical tests.
We are rebuilding the openSIM (structured illumination microscopy) around the small DLP2000 projector chip from TI's DLPDLCR2000EVM. A single-mode fibre (488 nm or 635 nm) lights the micromirror chip. The stripe pattern on the chip is relayed into the microscope, and a mask in the Fourier plane keeps only the light that forms the stripes.

What is new
- One printed block on an aluminium plate holds everything: fibre collimator, the DLP chip on a kinematic mount, the relay lenses and the Fourier mask on a small flexure stage.
- Swinging fibre arm: the collimator sits on a carriage that turns ±8° about the chip, so the illumination angle can be set on the bench for each colour.
- 1:1 relay with 50 mm lenses: the second lens hangs out of the block in a tube, which makes it easy to adjust.
- Electronics: the adapter board was redrawn in KiCad with parts the board maker can assemble.
Simulation
We model the projector chip as a tilted mirror plus a grating in Optiland and trace the light from the fibre into the microscope. Below: the beam through the collimator, the chip and the relay at 488 nm.

In the FRAME
The block goes into the upper layer of the FRAME's OPM. A tube lens images the Fourier plane into the back focal plane of the objective, as for the larger DLP projector (see that page).