Coordinates and axes

The FRAME has four motorised axes. X, Y and Z move the sample; A moves the objectives.

Axes of the FRAME

AxisWhat movesTravelDrive
Xsample plate, left and right as seen from the front127 mmNEMA 11 motor, 1 mm ball screw
Ysample plate, front and back86 mmNEMA 11 motor, 1 mm ball screw
Zsample plate, up and down (focus)8 mmtwo NEMA 8 actuators on wedge stages
Aobjective slider under the plate, along Xtwo positions, 29.8 mm apartNEMA 11 motor, 1 mm ball screw

The objectives never move up or down. Focus always moves the sample. Details: Mechanics.

Origin and offsets​

  • The stage zero corresponds to well A1 of a standard multiwell plate in the holder.
  • ImSwitch shows positions relative to that zero. The physical position differs by an offset per axis, stored in the configuration (stageOffsets); see Configuration (JSON).
  • The calibration target supplied with the unit has a centre cross at x = 63.5 mm, y = 46 mm in stage coordinates. Use it to check the offsets; see Calibrate stage.

Directions and homing​

  • The positive direction and the homing direction of each axis are set in the configuration (homeDirection, homeEndstoppolarity per axis). They can differ between units; check yours before you script moves.
  • X, Y and A home at start-up by default; Z does not. See Homing.
  • Units ID10–ID20 run without position feedback: after a crash or an emergency stop, home again before you trust a coordinate.

Entering coordinates​

Positions are entered in micrometres in ImSwitch and in the REST API. One microstep is 0.3125 µm on X, Y and A. See ImSwitch UI reference.